RI Detector RI-101 for High Performance Liquid Chromatography
Showa Denkos Shodex RI-101can be used with many HPLC systems.
- Refractive Index Range: 1,00 til 1,75
- Drift: 0,2micro-RIU/h
- Linearity range: 600micro-RIU
- Noise:2, 5 nRIU (response: 1,5sec)
- Response: 0,1; 0,25; 0,5; 1; 1,5; 2; 3; 6 sec
- Cell voume: 8micro-L
- Flow rate (usual): 0,2 til 3,0 mL/min
- Max back pressure: 50kPa
|
|

See also the RI 201H detector |
See also Versions RI-102 and RI 104
The Shodex RI-100 series includes versatile and highly sensitive RI detectors equippd with a color liquid crystal display, an automatic start-up function, and a validation wized, and can be used with various HPLC systems.
- Equipped with color liquid crystal display, the detector monitors status of chromatograph in real time.
- The automatic start-up function automates complicated operations such as the substituting eluent in the reference cell and checking the baseline stabilization.
- The validation wizard achieves component validation with ease.
- Optimization of the temperature control method shortens the time for stabilization after the start-up and improves baseline stavility.
- The leak sensor automatically stops the pump in case of solvent leakage.
- External input and output terminals and RS232C communication ports permit advances automation of the system.
|
|
 |
Specifications
| Model |
RI-101 |
RI-104 |
RI-102 |
| Analysis |
Semi-micro |
Preparative |
| Flow cell type |
2 chanber-type |
| Measuring method |
Deflection type |
| Reflactive Index range |
1.00 to 1.75 |
| Drift |
0.2micro-RIU/h |
0.2micro-RIU/h |
2micro-RIU/h |
| Measuring range |
0.25 to 512 micro-RIU |
0.25 to 512 micro-RIU |
2.5 to 5120 micro-RIU |
| Linearity range |
600micro-RIU |
600micro-RIU |
6000micro-RIU |
| Noise |
2.5nRIU (response: 1.5sec) |
5nRIU (response: 1.5sec) |
25nRIU (response: 1.5sec) |
| Response |
0.1 , 0.25 , 0.5 , 1 , 1.5 , 2 , 3 , 6 sec |
| Auto zero |
Full auto zero |
| Auto zero range |
All range |
| Off-set range |
0 to 500mV (Same as integrator output) |
| Off-set response |
10mV (Same as integrator output) |
Integrator output
(Sensitivity) |
DC 0 to 1V
(2mV/micro-RIU,
8mV/micro-RIU) |
DC 0 to 1V
(2mV/micro-RIU,
8mV/micro-RIU) |
DC 0 to 1V
(0.2mV/micro-RIU,
0.8mV/micro-RIU) |
| Cell volume |
8micro-L |
2.5micro-L |
8micro-L |
| Flow rate |
(Usual) |
0.2 to 3.0mL/min |
0.2 to 1.0mL/min |
1.0 to 50mL/min |
| (Max.) |
10mL/min (solvent: pure water) |
1.0mL/min (solvent: pure water) |
100mL/min (solvent: pure water) |
| Max. back pressure |
50kPa |
| Internal volume |
IN to Cell: ca. 60micro-L
Cell to OUT: ca. 600micro-L
All(Cell to OUT: ca.670micro-L |
IN to Cell: ca. 10micro-L
Cell to OUT: ca. 355micro-L
All(Cell to OUT: ca.370micro-L |
IN to Cell: ca. 120micro-L
Cell to OUT: ca. 510micro-L
All(Cell to OUT: ca.640micro-L |
| Recorder output |
0 to 10mV/FS |
| EXternal input |
Purge On/Off, Auto Zero, Marker |
| EXternal output |
(1) READY (Automatic start-up)
(2) LEAK
(3) ERROR (OVER HEAT/LOW LIGHT INTENSITY/
NULL GLASS HOME POSITION/LOST PARAMETERS/OPTICAL BALANCE)
(Cntact capacity : DC24V 0.1A max.) |
| Temp. control |
OFF, 30 to 50deg-C (1deg-C step), 77deg-C Temp. fuse |
| Communication port |
RS232C |
Operational support
functions |
(1) Autopamtic start-up (Start Up Sequence), (2) Span/Validation Guide,
(3) Real Time Baseline Monitor |
| Wetted materials |
Stainless steel 316, Teflon, Quarts glass |
Powersource,
Power consumption |
AC 100 to 240V±10%, 50/60Hz, 150VA max. |
| Demensions, Weight |
W260 x D400 x H200 (mm), ca.13kg |
| Accessories |
Power cable, single code, connector tube, fuse, operation manual |